Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Personne à contacter : Bruce Lu, +86 29 89284429
Bonne affaire : acheter au vendeur
Nous vous invitons à lire nos conditions générales d'utilisations. Vous pouvez aussi vous rendre sur nos FAQ et consulter notre page d'informations sur les risques liés à la contrefaçon.
|
Cette page concerne les importateurs et exportateurs de Normal Problem and Solution in GC Test Rechercher dans la catégorie : Divers Rechercher dans la catégorie : solution, problem, test, normal |
Wednesday 20 May 2015
AI101P Single Port KVM over IP The KVM over IP technology (IP-KVM) combines digital remote KVM access via IP networks with comprehensive and integrated system management. The IP-KVM defines a new class of remote KVM access devices that can save your money, time, space, and equipment...
ACAFA Information Co.
- epanio
- 335 - Dasi Township
- +88 6 33 07 13 00
Tuesday 08 September 2015
Module parameters HD-720P-38*38mm 1.0megapixel low-illumination HD web camera Network Camera - Sensor type 1/4 inch、1.0megapixel , CMOS low-illumination width dynamic sensor - Minimum illumination Color: 1 Lux/F1.2, black and white: 0.02Lux/F1.2 - Electronic shutter 1/5 second ~ 1/50...
ZT Technology Group Ltd.
- ellazhan
- 852 - Fo Tan
- +85 2 26 75 75 58
Tuesday 09 December 2025
▲VE-30RFM RX (Front / Rear) ▲VE-30RFM TX (Front / Rear) VE-30RFM HDMI to RF Matrix Extender is designed to convert HDMI signal to HD digital TV signal based on DVB-T CATV signal, it uses 64QAM mode to realize the HD digital TV signal conversion and transmit via coaxial cables for...
Meicheng Audio Video Co., Ltd.
- meicheng
- 23511 - New Taipei City
- 886 2 82280311
Paiement sécurisé